Customization: | Available |
---|---|
Certification: | ISO9001, CE |
Salt Rejection Rate: | 99.0% |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Technical Parameter
Model |
OK-MP10 |
OK-MP20 |
OK-MP40 |
OK-MP60 |
Output |
10L/h |
20L/h |
40L/h |
60L/h |
Feed water requirements |
Tap water,temperature 10~35ºC,pressure:0.15-0.35Mpa; TDS:≤200mg/L |
|||
RO water quality |
|
|||
Conductivity(25ºC) |
3-6μs/cm |
|||
Iron rejection rate |
99%(new RO membrane) |
|||
Organic rejection rate |
>99%,when MW>200Balton |
|||
Paeticles and bacteria rejection rate |
>99% |
|||
Ultrapure water quality |
|
|||
Resistivity(25ºC) |
18.2MOhm-cm |
|||
Bateria |
<1CFU/ml |
|||
Particle |
>0.22um/ml<1um/ml |
|||
AC requirement |
110/220V,50/60Hz |
|||
Dimension(L*W*H-mm) |
410*300*500mm |
|||
Weight(kgs) |
20kgs,30kgs,50kgs,70kgs |
|||
Standard configuration |
Main body,pressure tank |
The water purifying technology process is divided into source water pretreatment system,reverse osmosis system and ultra-pure system.
1. Source water pretreatment system: PP fiber filter + AC filter + Softener
2. Reverse osmosis system: It has high removal rate more than 99% for high valent ions, colloids, bacteria and organic matter (including endotoxin) with molecular weight greater than 300Dalton.
3. Ultra pure system: Ion exchange resin using H + exchange cation out the water in the filter ,using OH- exchange anions out the water , the replaced H + and OH- combined to produce H2O, ultra-purified column filling the United States Dow UPW-450 Ultra-pure water for nuclear grade resin, the water resistance of up to 18.2MΩ.cm.