Product name
|
Ethylene Oxide Tvoc Gas Chromatograph Analyzer
|
Model
|
OKGC-8960B
|
Relative humidity
|
< 85%
|
Power
|
2000W
|
Measuring range
|
customizable
|
Measuring object
|
customizable
|
Net Weight
|
40kg
|
Size
|
550*535*485mm
|
Oven size
|
240×250×155mm
|
Column installation interval size
|
152.4mm(6 inch standard interface)
|
Oven temperature control
|
Room temperature plus 8ºC~400ºC, (optionally set in 1ºC increments)
|
Lift rate
|
0.1ºC~50ºC (increment of 0.1)
|
Temperature fluctuation
|
Not more than ±0.1ºC; (Ambient temperature change 10ºC or power supply voltage change 10%)
|
Temperature gradient
|
±1% (temperature range 100ºC~350ºC)
|
Ambient temperature
|
5ºC~35ºC
|
Supply voltage
|
220±10%
|
Power supply frequency
|
50±0.5Hz
|
Sampling system
|
|
Maximum use temperature
|
400ºC
|
Number of inlets
|
Can be equipped with up to 3
|
Injection mode
|
Filled injection, split capillary injection, split/splitless capillary injection
|
Detection Systems
|
|
1.Hydrogen flame ion detector (FID)
|
|
Minimum detection limit
|
≤5 x 10-12g/s (n-hexadecane)
|
Dynamic range
|
≥10
|
Baseline noise
|
≤2 x 10-13 A
|
Baseline drift
|
≤5 x 10-13 A/30 min (after the instrument is stable for 1 hour)
|
Linear dynamic range
|
≥107
|
Insulation strength
|
≥1500 volts, 220V leakage current ≤ 1mA
|
2.Thermal conductivity detector (TCD)
|
|
Sensitivity
|
S≥8000mv.ml/mg (n-hexadecane)
|
Baseline noise
|
≤20uv
|
Baseline drift
|
≤100uv/30min (after the instrument is stable for 1 hour)
|
Linear dynamic range
|
≥104
|